GENERAL
RULES FOR TROUBLESHOOTING BY COMPONENT SUBSTITUTION
Always make sure
cable connectors are fully inserted into their mates and that the threaded
collars are tight
- The
most likely cause of a system malfunction is a defective tube head cable.
The cable alone may be defective or the defective cable may have
damaged other system components.
- If
a system is malfunctioning, never take a good control unit or tube head
from another working system and try it on the malfunctioning system
without first substituting a known good tube head cable.
- If
the system still does not work after substituting the good cable, do not
then assume the original cable is good; set it aside.
- After
insuring a good cable is installed, the next component to substitute is a
known good tube head for the questionable one.
- When
testing the system after substituting the good tube head, always start in
the Operate Mode with a 0kV and 0mA on the set line.
After pressing the x-ray ON, first advance the mA to 5.0mA and then
the kV one kV at a time. mA should be flowing and at 5.0 by the time 25kV
is reached. Continue to advance the kV up to 100kV.
If no problems are encountered up to 100kV, warm the tube head up
using the >30 days autowarm mode.
If problems occur at low kV do not keep trying to advance the kV.
- Do
not assume the original tube head is good.
- If
malfunctions still occur, substitute a good Control Unit.
Start in the Operating Mode and advance mA and kV as in Step 2A.
- The
system should be operating properly now if all of the components used for
substitution were good.
- Now
check out the original tube head by putting it back on the system.
Again start testing in the Operate Mode with 0kV and 0mA set in on
the control unit. After
pressing the x-ray ON, advance the mA to 5mA and the kV one kV at a time.
mA should be flowing and at 5,0 when 25kV is reached.
Continue to advance the kV up to 100KV. If problems occur at low kV, do not keep trying to
advance the kV. If no
problems are encountered up to 100KV, auto-warm the tube head at the
>30 day rate.
- Continuity
test cables pin on one end to corresponding pin on other end and from each
pin to all the other pins and the shell on the same end of the cable.
BEWARE: a cable may be good laying in one position but be defective
in another position.
PROBLEM DIAGNOSIS SUGGESTIONS
- Repeated
Are Detect Message
- Low
Line Voltage - Line voltage below 100 volts (continuously or
intermittently) causes the x-ray generation to stop and the ARC DETECTED
message to be displayed. Measure
line voltage under lead. Use
power source greater than 100 volts.
Don't use extension cord.
- Insufficient
Warm Up - Tube reactions can result from inadequate warm-up causing the
ARC DETECTED message to be displayed.
Manually warm up using the procedure described in NOTES on Page 4-6
of T.0 33B33-31-11.
- Defective
Components - Return to Depot for repair.
- Interlock
Flow Message
- Interlock
Open
i.
Interlock jumper missing
ii.
Defective cable connecting external interlock/warning system
iii.
Defective external interlock system
- Defective
Control Unit - A defective or incompatible external interlock system can
damage the Control Unit by shorting pins A or B to ground or by feeding
external voltages into Pins A or B. Check for shorts to ground on Pins A
and B of the cable and also for the presence of voltage on Pins A and B of
the cable. Try disconnecting
interlock cable and using supplied interlock jumper to isolate problem.
- Insufficient
Coolant Flow
i.
Check and clean filter in cooler
ii.
Check for clogged hoses or improperly mating hose fittings
iii.
Continuity check cooler cable
iv.
Bad flow switch in cooler - replace flow switch
- Circuit
Breaker Trips
- Trips
immediately on application
i.
Line voltage in excess of 130 volts - measure line voltage.
Use power source, less than 130 volts.
- Trips
after one to two second delay
i.
Short in cooler or cooler cable. Substitute different cooler or cooler cable.
Continuity test cable.
ii.
Defective control unit
iii.
Intermittently high line voltage in excess of 130 volts
- Trips
when x-ray ON switch is pressed.
i.
Defective control, tube head, or tube head cable.
Continuity test cable.
- Temperature/Pressure
Message
- Low
tube head gas pressure or high anode temperature - See T.O. 33B3-3-31-11.
- Defective
tube head cable - Continuity test cable or substitute with known good
cable.
- Loose
or Defective Tube Head Anode Thermostat - Separate two halves of small
white connector attacked to wire leading out of tube head pressure switch
and down the side of the tube head. Measure
the resistance between the connector half on the wire leading down the
side of the tube head and the bare 10-32 screw head on the tube head end
plate. The resistance should
be less than 5 ohms. Look for
breaks in the wire, tighten or replace thermostat.
- Defective
Tube Head Pressure Switch - Continuity test pressure switch by measuring
resistance between half of white connector that is attached to pressure
switch and Pin A of tube head end plate connector.
Resistance should be less than 3 ohms.
- Repeated
Excess kV Message
- Defective
tube head cable, control unit or tube head. Substitute good components to isolate problem.
- X-Ray
Switch Error Message
- Message
appears within 3 seconds of switching the mains on.
X-ray ON switch is shorted. Repair
or replace.
- Message
appears while system is generating x-rays.
i.
Insufficient warm-up. Perform
manual warm up –See notes on Page 4-6 of T.O.33B3-3-31-11.
ii.
If re-occurs repeatedly, tube head cable or tube-head may be,
defective.
- Interrupted
Restart Message
- Message
appears after fault condition has been corrected.
- May
appear after intermittent fault such as erratic coolant flow, loose
thermostat, temperature/pressure conditions, etc.
- Black
Blocks on LCD Displays
- These
black blocks are caused by an electrical pulse from a tube reaction
interfering with the digital display.
Turn the unit off and then back on and warm up at slowest rate.
- If black blocks repeatedly occur, the tube head
cable or the tube head may be defective.